    Netherland evaluation    

 Synthesis of the evaluation report of the Sterinis® process for the disinfection of units of care at Gooi-Noord Regional Hospital (Netherland)

 

The evaluation of Sterinis® at the Gooi-Noord regional hospital (Netherlands) was carried out on a room presenting an architecture making difficult airborne disinfection.


The selected stock "Staphylococcus aureus" is of clinical origin. The 15 samples points of 10 cm² each one are distributed in all the room, the inoculum is of 105cfu/ml for each point.


The results are presented for each situation. The table presents for each point of measurement, the nature of the sample point and the results of the sample after 30 minutes of contact then 1h30 of time of contact (then after 20h of incubation). Only a point does not profit from a reduction of 5log.

B14 Room 16

1st  series (contact :time 30 mn)

2nd series : (contact: time 1h30)
 treatment 20 houres after the first treatment

name place of the sample

Sample N#
( 1 to 15)

Samples at
105 cfu / ml
(S. aureus)

Sampling directly on the site

Sample N#
( 1 to 15)

Samples at
105 cfu / ml
(S. aureus)

Sampling directly on the site

Cabinet N#1

1

0 - growth

0 - growth

1

0 - growth

0 - growth

Cabinet N#2

2

0 - growth

 

2

0 - growth

 

Light Element

3

S. Aureus

 

3

0 - growth

 

Window sill

4

0 - growth

0 - growth

4

0 - growth

0 - growth

Radiator

5

0 - growth

 

5

0 - growth

 

Bed N#2

6

0 - growth

0 - growth

6

0 - growth

0 - growth

Matelat N#2

7

0 - growth

 

7

0 - growth

 

Téléphone

8

0 - growth

 

8

0 - growth

 

TV

9

0 - growth

0 - growth

9

0 - growth

0 - growth

Tablette N#1

10

0 - growth

 

10

0 - growth

 

Tablette N#2

11

0 - growth

 

11

0 - growth

 

Cabinet bathroom

12

0 - growth

0 - growth

12

0 - growth

0 - growth

Wash bin toilet

13

0 - growth

0 - growth

13

0 - growth

0 - growth

Shower

14

0 - growth

 

14

0 - growth

 

Bed N#1

15

0 - growth

 

15

0 - growth

 


The results highlight the principal following points:

Sterinis® is a device making possible to diffuse a disinfecting dry mist in all the volume of the room, even in not directly accessible areas


Sterinis® is able to apply a reduction of 4,8log by airborne disinfection as early as 30 minutes of contact in all the part to a clinical stock of “Staphylococcus aureus


  • Sterinis® is a device making possible to diffuse a disinfecting dry mist in all the volume of the room, even in not directly accessible areas.

  • Sterinis® is able to apply a global reduction of 4,8log by airborne disinfection as early as 30 minutes of contact in all the part to a clinical stock of “Staphylococcus aureus.

 Evaluation report's Doctors C.P. Timmerman et R.J.D. Samson: Full evaluation report